Abstract
A practical and reproducible high-performance liquid chromatographic method using normal solid-phase extraction has been developed for the simultaneous analysis of twelve non-steroidal anti-inflammatory drugs (NSAIDs) in human urine. A urine specimen mixed with acetate buffer pH 5.0 was purified by solid-phase extraction on a Sep-Pak Silica cartridge. The analyte was chromatographed by a reversed-phase Inertsil ODS-2 column using a phosphate buffer-acetonitrile at pH 5.0 as the mobile phase, and the effluent from the column was monitored at 230 or 320 nm. Absolute recoveries were greater than 73% for all of the twelve NSAIDs. The present method enabled simple manipulation and isocratic HPLC with UV analysis as well as high sensitivity of 0.005 microg/ml for naproxen, and 0.05 microg/ml for sulindac, piroxicam, loxoprofen, ketoprofen, felbinac, fenbufen, flurbiprofen, diclofenac, ibuprofen and mefenamic acid as the quantitation limit in human urine using indomethacin as an internal standard.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.