Abstract

The temporal resolution uniformity of a time-dilation framing camera is studied, and the ideal photocathode (PC) pulse curve is determined, while the temporal magnification factor is kept constant. To obtain the ideal curve, a series of linear pulses with the same slope are superposed. The variance in the temporal resolution and the number of linear pulses required are compared, while the superposition results with different slopes are used as the PC voltage. As the slope of the linear pulses decreases, the variance decreases, which means that the uniformity of the temporal resolution is improved, but the number of linear pulses required increases. In this study, linear pulses with a slope of 1V/ps are superposed. Nine linear pulses with a front edge time of 200ps, amplitude of 200V, and flat top time of 1ns are superimposed to approximate the ideal PC pulse curve with a constant temporal magnification factor of 10; the trigger times of the pulses are 0, 0, 0, 185, 200, 350, 450, 605, and 790ps. When the superposition result is applied as the PC voltage and the measured signal is synchronized to the PC pulse at 128ps-1ns, the temporal resolution error is within 5%.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call