Abstract

The in vivo monitoring Lab at KIT uses two phoswich detectors for routine lung counting. A simplified model of one of them has been implemented in the two Monte Carlo codes EGSnrc and MCNPX. The active part of the detector consists of a crystal of NaI(Tl) and one of CsI(Tl): the energy deposited in both the crystals have been studied to consider the effect of the anticoincidence logic, present in the read-out electronics of the detectors and not yet studied with Monte Carlo simulations. Only the NaI(Tl) crystal has then been used to study the escape peaks at several energies, which are more prominent at low energies. The results from the two codes have been compared. The comparison of the codes predictions for the escape peaks has been then extended using the model of a germanium detector.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.