Abstract
Using Monte Carlo simulations we optimized the wavelength and source-detector distance (SDD) of a reflectance-based spectroscopic device used for measuring subcutaneous fat thickness. As the optical properties of muscle, fat and dermis are wavelength dependent, it is necessary to choose a wavelength that is highly sensitive to fat but insensitive to water and melanin. The SDD is important since it determines average photon penetration depth. With a tissue optics plug-in for the GEANT4/GAMOS system and published ex vivo tissue optical properties we were able to predict the behavior of different device configurations when used with varying thicknesses of fat, melanin concentrations or hydration levels. Our results indicate that the ideal wavelengths for fat measurement are 630-650 nm with an SDD of 2.6-29 cm. We also examined the potential of using near infrared (NIR) spectroscopy to determine tissue hydration levels, but concluded that this wavelength range was not ideal.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.