Abstract

A new personal dosimeter for neutrons, using three TLD 600s and one TLD 700 in a cadmium housing and detecting both incoming and albedo neutrons, has been designed, developed, tested and simulated by the code MORSE. Its response in the energy region between 10 −8 and 10 MeV is more similar to ICRP fluence-dose equivalent calibration factor behaviour with respect to previous albedo dosimeters, and particularly fast neutron tissue dose equivalents are less underestimated. Present data confirm that both accuracy and precision fulfil the International Commissions requests (NCR, NCRP, ANSI). Theory, simulation and experimental results obtained with a laboratory prototype are presented and discussed.