Abstract
A rapid, selective and sensitive method for the determination of dihydroergotamine (DHE) in serum was developed. Dihydroergocristine (DHEC) was used as an internal standard. Human and rabbit serum samples were extracted using commercial solid-phase cyano (CN) columns. Proteins were washed from these columns with pure acetonitrile, resulting in clean extracts. Extracts were subsequently separated by HPLC in an isocratic way, using a reversed-phase C 18 analytical column. Fluorometric detection was performed at excitation and emission wavelengths of 277 and 348 nm, respectively. Calibration curves with amounts of DHE ranging from 2 to 32 ng, were linear. The limit of detection found for DHE was 0.2 ng, extracted from 0.5 ml rabbit or from 2.5 ml human serum. The limit of quantification in serum of both species was 0.7 ng. The method has been shown to be suitable for monitoring DHE in serum during pharmacokinetic studies in rabbits.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.