Abstract

A simple, non-moving dual-stage CO2 jet modulator is described, which cools two short sections of the front end of the second-dimension column of a comprehensive two-dimensional gas chromatograph. A stream of expanding CO2 is sprayed directly onto this capillary column to trap small fractions eluting from the first-dimension column. Remobilization of the trapped analytes is performed by direct heating by the GC oven air. Installation, maintenance and control of the modulator is simple. Focusing and remobilization of the fractions is a very efficient process, as the bandwidths of the re-injected pulses are less than 10 ms. As a result, alkane peaks eluting from the second-dimension column have peakwidths at the baseline of only 120 ms.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.