Abstract

A simple interface has been developed to couple capillary electrophoresis (CE) to inductively coupled plasma atomic emission spectrometry (ICP-AES) for metal speciation. A concentric glass nebulizer with elongated tip is used as the CE–ICP interface. The CE capillary is the central tube of the nebulizer. A platinum wire is wrapped across the exit end of the CE capillary to provide electrical connection to the CE power supply. No sheath flow of buffer solution is needed. A simple cooling system has also been developed. A peristaltic pump circulates water through a plastic tube that encloses the section of the CE capillary between the CE instrument and the ICP spectrometer. Characteristics of the CE–ICP interface, e.g., elution time, nebulization and transport efficiency and peak broadening, versus carrier gas flow-rate have been studied. Comparisons to a previous design with the Pt electrode inserted into the end of the CE capillary are made where appropriate. The reproducibility (RSD) in ICP emission intensity of the system is <4%. Detection limits of Cr and Cu are ∼5 ng/ml.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.