Abstract

A modified concentric nebulizer was used as the interface to couple capillary electrophoresis (CE) to inductively coupled plasma atomic emission spectrometry (ICP-AES). The CE capillary replaces the central tube of the concentric nebulizer. The tip of the nebulizer tapers slowly to allow uncertainty in the position of the capillary. A platinum wire was inserted into the CE capillary to provide electrical connection to the CE power supply. pH changes inside the capillary due to electrolysis of the background buffer electrolyte was small and has minimal effects on the CE separation. The peak broadening effects due to the nebulizing gas flow, however, were significant. Resolution decreases quickly when the flow-rate of the carrier gas increases. Sample stacking technique was used to improve the resolution of species of opposite charge, e.g., Cr(VI) vs. Cr(III) ions. Detection limit of Cr based on peak area is ∼10 ppb for the CE–ICP-AES system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.