Abstract
Using simple experimental models, we evaluated the generation, configuration and echo intensity of vertical artifacts by varying the point or plane of contact and height of objects that correspond to sources of vertical artifacts in the subpleural space. We used an ultrasound gel spot to imitate the source and a block of bacon as a chest wall phantom. As the size of the point of contact between the gel spot on the polypropylene sheet and the phantom decreased by peeling the sheet, a vertical artifact measuring ≤1 cm was generated and/or extended deeper, finally reaching 10 cm in depth. Next, objects of different shapes made using gel balls were used to observe the generation of artifacts and measure and compare the echo intensity. For a given shape, the intensity was markedly higher in one model with the point of contact than in the other model with the plane of contact. With the same point or plane of contact, the echo intensity was higher in the taller model. The size of the point or plane of contact and height of the source were observed to be key factors in the generation, length and echo intensity of the artifacts.