Abstract

In a pulse-height analysis system using a gated periodic pulse train for the correction of counting losses, it is shown that both the dead-time and pile-up losses can be corrected for if the clock pulses have the same width as the so called true pile-up time of the pulse processing system, and the clock pulses are rejected from the pulse train if a coincidence between the shaped clock pulses and the busy time of the system is detected. The technique was tested both with random-pulse generators and a Ge(Li) detector ( 60Co and 137Cs sources). At counting rate up to 50 kHz, it is shown that the correction was accurate to ± 1%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.