Abstract

A method was proposed for the direct determination of carbon dioxide in air using a sorption tube specially constructed for this purpose. The method is based on measuring the volume of air consumed for the neutralization of a known amount of barium oxide hydrate in the tube by carbon dioxide contained in air. The equivalence point is detected by the minimum electrical conductance of the barium hydroxide solution. To verify the accuracy of the method, a technique based on the use of a calibration mixture of air with the CO2 concentration varying by the inverse exponential law was developed. The method is suitable for the determination of CO2 in air in a wide concentration range and for the verification and calibration of indirect methods for the determination of CO2. With an insignificant decrease in accuracy, the equivalence point can be detected by the change in the color of phenolphthalein. This significantly simplifies the method and allows its wide use in practice, e.g., in laboratory works in environmental science or for demonstration purposes in educational organizations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.