Abstract

Mass transfer from the first-dimension to the second-dimension column in comprehensive two-dimensional gas chromatography (GC × GC) is normally quantitative, which means that the peak areas produced by GC × GC are equal to those obtained in one-dimensional gas chromatography (GC). In view of this relationship, it was investigated whether normal GC calibration could be used to quantify compounds analysed by GC × GC. Quantification of polycyclic aromatic hydrocarbons (PAHs) in sediment was used as a test case. It was concluded that more laborious and time-consuming GC × GC area calibration can be replaced by GC area calibration if separation quality and quantity prerequisites are met.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.