Abstract
A simple, selective and sensitive heavy atom-induced room temperature phosphorimetric method (HAI-RTP) is described for the determination of naproxen (NAP) in pharmaceutical preparations. The phosphorescence signals are a consequence of intermolecular protection when analytes are, exclusively, in presence of a heavy atom salt and sodium sulfite as an oxygen scavenger to minimize RTP quenching. These variables selection constitute the basis of a HAI-RTP method for the determination of naproxen (detection limit 17.6 ng ml −1; 1.71% relative standard deviation at 250 ng ml −1). The method has been applied satisfactorily to the analysis of pharmaceutical preparations.
Paper version not known (Free)