Abstract
ABSTRACT Silver ion-selective electrodes based on 2-mercaptobenzimidazole (MBI) and 2-mercaptobenzothiazole (MBT) carriers have been developed. The poly(vinyl chloride) (PVC)-based membranes containing MBI or MBT, plasticizer and sodium tetraphenylborate, as anion excluder and membrane modifier, were directly coated on the surface of graphite electrodes. The electrodes exhibited near-Nernstian slopes of 60.2 and 57.8 mV per decade of Ag+ concentration for MBI and MBT, respectively. Calibration plots were linear over a wide concentration range of 1 × 10−6–1 × 10−2 M Ag+. The electrodes have shown detection limits of 6.3 × 10−7 and 4 × 10−7 M for MBI- and MBT-based electrodes, respectively, and revealed good selectivities over a number of alkali, alkaline earth, transition and heavy metal ions. The electrodes have response times between 10–20 s and can be used for at least two months without a considerable divergence in their potential responses Application of the electrodes for potentiometric titrations and determination of Ag+ ion concentration in photographic waste solutions are reported.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.