Abstract

Electrophoretic separations conventionally rely on chromogenic, fluorogenic, or redox properties for analyte detection that, in many instances, involve chemical modification of samples prior to analysis. For analytes natively lacking chemical signatures, refractive index-based measurements are appealing as a method to detect these molecules without pretreatment. Microring resonators are a type of whispering gallery mode sensor capable of detecting bulk changes in refractive index. Here, we demonstrate the use of silicon photonic microring resonator arrays as a postcolumn detector for capillary electrophoresis. In this approach, we establish the universal detection capabilities of microrings through calibration with analytes lacking unique spectral signatures. Separations of small molecule mixtures are demonstrated using capillary zone electrophoresis. For these separations, the microring resonators maintain a linear response over several orders of magnitude in concentration for three candidate small molecules. Successful separation of three sugars with direct detection is also demonstrated. We further present the successful separation and detection of three model proteins, exemplifying the promise of microring resonators arrays as a biocompatible detector for capillary electrophoresis. Additionally, the spatially offset, array-based nature of the sensing platform enables real-time analysis of analyte mobility and performance characterization-a combination that is not typically provided using single-point detectors.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call