Abstract

A large area silicon drift detectors (SDDs) system and its readout electronics have been developed by the SIDDHARTA-2 Collaboration, aiming to perform high-precision light kaonic atoms x-ray spectroscopy for the investigation of the strong interaction in the low-energy QCD regime. To perform these measurements, a linear energy response and a good energy resolution are mandatory requirements for the system, to be preserved along the whole DAQ (analog and digital) chain; such task is made even harder in the experimental environment of particles colliders, where the high background due to ionizing particles and radiation is present. The energy response of the SDDs system has been characterized with the beam-originating background generated during the commissioning phase of the DAΦNE electron–positron collider (INFN-LNF) in early 2020. The data analysis has been optimized to describe the system’s response and the background. The calibration procedure demonstrates that, despite the high and variable background of the collider, the energy response of the system is linear at the level of few eV (), with an energy resolution of eV for the Fe K line.