Abstract
The signal-to-noise ratio (SNR) in magnetic resonance imaging represents one of the system operating variables that must be determined both for evaluating the performance of different imaging protocols on a particular machine, and for monitoring machine performance as part of a routine quality control (QC) program. Utilizing a phantom and set of automated analysis programs currently under development, this study evaluated several ways of measuring image signal and noise and demonstrated the importance of utilizing measured voxel volumes as opposed to nominal volumes in the calculation of SNR. The NEMA proposed standard for SNR is compared with several other SNR measures and is recommended as the measure to be used in routine SNR reporting. The importance of utilizing other SNR measures in addition to the NEMA proposed standard for routine QC is discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.