Abstract
Objective: The objective of this study was to compare the signal-to-noise ratios (SNRs) of 6 intraoral direct digital sensors. The sensors were Visualix 1, Visualix 2, DenOptix, Computed Dental Radiography (CDR), CDR APS, and Digora. Study Design: The sensors were exposed by using a dental x-ray machine operating at 50 kV, 8 mA and various exposure times. The phantom used was a hole measuring 1 mm in diameter and 1.25 mm in depth, drilled in a 10-mm–thick aluminum block. Digital images were transferred as 8-bit tagged image file format files and analyzed by using a personal computer. Results And Conclusions: SNR values increased up to a maximum value at a certain exposure level and then, depending on the system, either decreased or maintained a high level with increased exposure. The Digora system produced the highest SNR values over a large exposure range, whereas the Visualix 2 and DenOptix systems produced the lowest SNR values. The 2 CDR sensors produced rather high SNRs but over a very small exposure range.(Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2001;91:611-5)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.