Abstract

Ultraviolet (UV)-C irradiation is a promising method for microbial eradication on surfaces. Major developments have taken place in UV-C light-emitting diodes (LEDs) technology. In this study, we examined the suitability of UV-C LED-based surface disinfection in hospitals. We tested the efficacy of UV-C LED surface treatment on different microorganisms dried on a carrier surface or in a liquid solution. The influences of soiling, shading, surface material, radiation wavelength, microbial load and species on the disinfection performance were investigated. UV-C LED caused a reduction of >5 log10 levels of E. coli, S. aureus and C. albicans, whereas 3 log10 reduction was observed for G. stearothermophilus spores. The components of the medium led to a reduced UV-C LED efficiency compared to buffered solutions. We observed that the microbial load and the roughness of the carrier surface had a major influence on the UV-C LED disinfection efficiencies, whereas shading had no impact on inactivation. This study showed that UV-C is suitable for surface disinfection, but only under certain conditions. We showed that the main factors influencing microbial inactivation through UV-C light (e.g., intrinsic and extrinsic factors) had a similar impact when using a UV-C LED radiation source compared to a conventional UV-C lamp. However, the potential of LEDs is contributed by their adjustable wavelength and customizable geometry for the decontamination of medical devices and surfaces, and thereby their ability to overcome shading effects.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call