Abstract

Neurofilament light chain (NfL) has emerged as a biomarker of neuroaxonal damage in several neurologic conditions. With increasing availability of fourth-generation immunoassays detecting NfL in blood, aspects of pre-analytical stability of this biomarker remain unanswered. This study investigated NfL concentrations in serum and plasma samples of 32 patients with neurological diagnoses using state of the art Simoa technology. We tested the effect of delayed freezing of up to 7 days and statistically determined stability and validity of measured concentrations. We found concentrations of NfL in serum and plasma to remain stable at room temperature when processing of samples is delayed up to 7 days (serum: mean absolute difference 0.9 pg/mL, intraindividual variation 1.2%; plasma: mean absolute difference 0.5 pg/mL, intraindividual variation 1.3%). Consistency of these results was nearly perfect for serum and excellent for plasma (intraclass correlation coefficients 0.99 and 0.94, respectively). In conclusion, the soluble serum and plasma NfL concentration remains stable when unprocessed blood samples are stored up to 7 days at room temperature. This information is essential for ensuring reliable study protocols, for example, when shipment of fresh samples is needed.

Highlights

  • Neurofilament light chain (NfL) has emerged as a biomarker of neuroaxonal damage in several neurologic conditions

  • The overall mean sNfL concentration was 14.5 pg/mL, the mean pNfL concentration was 12.5 pg/mL

  • This study investigated a potential limitation for NfL research being constrained to immediate availability and accessibility of standardized sample processing, i.e., a biobanking procedure

Read more

Summary

Introduction

Neurofilament light chain (NfL) has emerged as a biomarker of neuroaxonal damage in several neurologic conditions. We found concentrations of NfL in serum and plasma to remain stable at room temperature when processing of samples is delayed up to 7 days (serum: mean absolute difference 0.9 pg/mL, intraindividual variation 1.2%; plasma: mean absolute difference 0.5 pg/mL, intraindividual variation 1.3%). The soluble serum and plasma NfL concentration remains stable when unprocessed blood samples are stored up to 7 days at room temperature. This information is essential for ensuring reliable study protocols, for example, when shipment of fresh samples is needed. There is no investigation with the primary hypothesis of confirming stability of NfL up to 7 days in unprocessed samples of both serum and plasma

Methods
Results
Conclusion