Abstract

A method is described for the quantitative multi-element analysis of small fragments (200–500 µg) of glass using an inductively coupled radiofrequency argon plasma source. The glass samples are digested with a mixture of hydrofluoric and hydrochloric acids and chromium is added as an internal standard. An ultrasonic nebuliser is used in order to reduce to a minimum the volume of solution required for each analysis. A single monochromator and detection system is employed, and the wavelength regions of interest are examined sequentially by means of a specially constructed control unit. The results for aluminium, barium, iron, magnesium and manganese show that the analysis of glass fragments in the range 200–500 µg can be achieved with coefficients of variation of approximately 10%. Standard glasses were analysed to assess the accuracy of the method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.