Abstract

Quantitative chemical analysis is a powerful tool for the classification and discrimination of small glass fragments in forensic and archaeometric analytical issues. In the choice of the analytical method the small dimensions and the irregular shape of the fragments must be accounted for. Several accurate and highly sensitive instrumental techniques successfully employed for the chemical analysis of small glass samples (AAS, ICP-AES and FAAS) are destructive, and therefore unacceptable for these applications. X-ray analytical techniques such as WDXRF, EPMA and SEM/EDX provide means for rapid, non destructive and sensitive multi-elemental analysis of glass fragments. In this study the analytical sensitivity and accuracy of WDXRF, EPMA and SEM/EDX for the analysis of small commercial glass fragments are investigated. The results are compared and discussed in order to establish a relation between the minimum size of glass samples and the accuracy and sensitivity of the analytical methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.