Abstract

The detection of human bone morphogenic protein-7 (BMP-7) was achieved using a sequential injection immunoassay (SIIA) system. The SIIA system is based on the binding between BMP-7 and anti-human BMP-7 (AbBMP7)–CdSe/ZnS quantum dot (QD) conjugates immobilized onto a glass disk or an optical fiber, using fluorescence detection at excitation and emission wavelengths of 470nm and 580nm, respectively. The AbBMP7–QD conjugates were prepared by conjugating anti-human BMP-7 antibody (AbBMP7) to hydrophilic CdSe/ZnS core/shell quantum dots (QDs). The SIIA system was fully automated using software written in the LabVIEW™ development environment. The analytical performance of the SIIA system was characterized with a number of variables such as carrier flow rate and elution buffer. Under partially optimized operating conditions, the SIIA system had a linear calibration graph at up to 10.0ngmL−1 BMP-7 (R2≥0.975) and a sample frequency of two samples per hour. The SIIA system with an optical fiber immunosensor was used to detect and quantify BMP-7 in spiked real samples obtained from a biological process with recoveries in the range of 95–102%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.