Abstract

A flow injection (FI) system is described for the sequential determination of periodate and iodate based on their reaction with iodide at pH 3.5. Two sample plugs were injected into the same carrier stream sequentially. One injection is for the iodate determination and the other for the sum of iodate and periodate determination. For iodate determination, molybdate solution buffered at pH of 3.5 was used for selective masking of periodate. The influences of reagent concentrations were studied by a univariable method and the influence of FI manifolds was studied using univariable and simplex method. Periodate and iodate can be determined in the range of 0.050–5.0 and 0.050–10 μg/ml, respectively. The 3σ limit of detection was 0.030 and 0.050 μg/ml for periodate and iodate, respectively. The proposed method has been applied for the sequential determinations of periodate and iodate in water samples.