Abstract

A multi‐element flow injection (FI) system with spectrophotometric detection was developed for sequential determination of Zn(II), Cu(II), Cd(II), Pb(II), and Hg(II). Dithizone reagent was used for the previously mentioned analytes and it was prepared on line by using a solid reagent column (SRC). Bear in mind that the most hazardous elements for environmental contamination are lead, cadmium, and mercury; a multi‐element FI system with preconcentration steps is proposed to determine their trace levels in natural waters. Results obtained for different samples agreed satisfactorily with those obtained by the reference method inductively coupled plasma‐atomic emission spectrometry (ICP‐AES). Detection limits were 5.4, 5.0 and 14 µg L−1 for Cd(II), Pb(II), and Hg(II), respectively. The percent of relative standard deviation (RSD) was better than 4.5% for each one.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call