Abstract
A multi‐element flow injection (FI) system with spectrophotometric detection was developed for sequential determination of Zn(II), Cu(II), Cd(II), Pb(II), and Hg(II). Dithizone reagent was used for the previously mentioned analytes and it was prepared on line by using a solid reagent column (SRC). Bear in mind that the most hazardous elements for environmental contamination are lead, cadmium, and mercury; a multi‐element FI system with preconcentration steps is proposed to determine their trace levels in natural waters. Results obtained for different samples agreed satisfactorily with those obtained by the reference method inductively coupled plasma‐atomic emission spectrometry (ICP‐AES). Detection limits were 5.4, 5.0 and 14 µg L−1 for Cd(II), Pb(II), and Hg(II), respectively. The percent of relative standard deviation (RSD) was better than 4.5% for each one.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.