Abstract
This paper is the first report on the retention behavior of synthetic oligonucleotides and nucleotide oligomers on a continuous-bedmatrix, strong-anion-exchange column. The separation mechanism is predominantly an anion-exchange process, but hydrophobic interaction plays a role as well. The separation is based on the chain length of the oligonucleotide. Both the addition of organic mobile phase modifiers and changes in column temperature affect the retention of oligomers significantly. A volatile buffer system (e.g., triethylamine acetate) could be employed to purify oligonucleotides, and no desalting procedure would be required after the column separation step. The recoveries from the separation are 70% or higher. The maximum loading capacity of an analytical column (35 x 7-mm i.d.) was found to be more than 366 micrograms.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.