Abstract

In this research study, a simple, rapid, sensitive, and cost-effective Reverse Phase High Performance Liquid Chromatography method was developed and validated for determination of Cetirizine and Levocetirizine in human plasma. The drugs were separated on an ACE Generix100-5 C18 RP (250 × 4.6mm, 5μm) column, using Acetonitrile and Distilled water pH7 (59:41v/v) as mobile phase with a flow rate of 1.5mL/min, retention time of 3.39 ± 0.02min, detection wavelength of 230nm, and column oven temperature at 26°C. The drug was extracted from human plasma using Methanol, the percent recovery was 97%. The LOD and LOQ for Cetirizine and Levocetirizine were 0.5ng and 1ng. The method was linear over the concentration range of 0.5-1000ng /mL for Cetirizine and Levocetirizine. The method was applied for analysis of drugs in plasma following oral administration of Cetirizine (10mg) and Levocetirizine (5mg) in smokers and nonsmokers of Pashtun ethnicity belonging from different districts of Khyber Pakhtunkhwa, Pakistan. The difference between Cetirizine plasma levels in smokers and non-smokers was significant with p -value 0.0279.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.