Abstract

The sensitivity enhancement of wavelength modulation surface plasmon resonance (SPR) biosensor was described by improving the baseline solution that made the resonant wavelength move to longer wavelength. The baseline solutions here used are ethanol solutions of 1.0%, 2.0%, 4.0% and 6.0%. The higher is the concentration of the ethanol solution, the longer is the resonant wavelength, and the lower is the concentration of the analyte that could be detected. The heat shock protein (Hsp) 70 are studied and when the baseline solutions used are 1.0%, 2.0%, 4.0% and 6.0% ethanol solutions, the lowest concentrations of Hsp 70 that can be detected are 1.00, 0.50, 0.20 and 0.10 μg/ml, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.