Abstract

We developed new measurement configurations based on organic electrochemical transistors (OECTs). Three types of COOH-functionalized bioreceptor layers were deposited on indium tin oxide (ITO) electrodes on poly(ethylene terephthalate) (PET) substrates and their performance was tested using single gate functionalization organic electrochemical transistor (S-OECT) and dual gate functionalization organic electrochemical transistor (D-OECT) configurations. The three layers included one p-type semiconductor, one insulator, and one self-assembled layer, and the dual gates were connected in series through buffer solutions, so the solution-electrode interfaces had the opposite polarities. We investigated the sensitivities of these systems using the human IgG antigen-human IgG antibody receptor pair for main experiments, and drifts of antibody-functionalized gates without analytes as control experiments. Drifts without analyte can obscure the real sensitivity. We show that the D-OECT has the capability to cancel the drifts, and is also beneficial for showing the sensitivity more exactly. This configuration has the ability to increase the accuracy of antibody-antigen interaction detection, and further decrease or eliminate the effect of ions in the buffer solution. We also prove that the D-OECT can work well with different bioreceptor materials, which indicates that the system can be further applied to different conditions.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call