Abstract

A sensitive and selective ion-pair reversed-phase HPLC method has been developed for the quantitative determination of KC 12291 and its major metabolite, KC 13194, in rat plasma and urine. An Ultrasphere ODS column constructed by using a mobile phase of 1 m M 1-octanesulfonic acid containing acetonitrile–0.1 M triethylamine phosphate buffer, pH 2.2 (29:71, v/v in plasma and 27:73, v/v in urine), an internal standard and a fluorescent detector (excitation 265 nm, emission 370 nm) were used for the separation and the quantitative determination, respectively. The plasma samples were made alkaline and both compounds were cleaned up by the use of liquid–liquid extraction. The limit of quantification was 10 ng/ml for KC 12291 in plasma and urine and for KC 13194 and 100 ng/ml in plasma, respectively. The assay has been validated with respect to system suitability, accuracy, precision, recovery, stability and ruggedness. All validated parameters were found to be within the necessary limits.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.