Abstract

A method has been developed for the rapid quantitative analysis of chloroquine and its metabolite desethyl-chloroquine in plasma, blood and urine using high-performance liquid chromatography. An ethylene dichloride extract of the alkalinized biological samples was extracted with dilute acid and chromatographed on a reversed-phase column. Phosphate buffer in acetonitrile was used as the mobile phase with perchlorate as the counter-ion. Ultraviolet absorption at 254 or 340 nm or fluorescence detection was used. The fluorescence spectra and the fluorescence quantum yield of the substances were determined. Chloroquine and desethyl-chloroquine concentrations in the range of 10 nmol/l (UV-detection) and of 0.5 nmol/l (fluorescence detection) could be accurately measured with a relative standard deviation of 12%. The method should be adequate for therapeutic and pharmacokinetic studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.