Abstract

Continuous flow plumbane generation for sample introduction into an inductively coupled plasma (ICP) and further determination of lead by atomic emission spectrometry (AES) has been investigated in order to increase the detection limits of conventional nebulization ICP-AES. Continuous hydride generation from different media is discussed and the generation of plumbane using potassium dichromate with lactic acid has been selected. Sensitivity, selectivity and accuracy of the corresponding determination of lead by ICP-AES are investigated. The proposed method (detection limit, 2 ng ml–1, precision 1.3% at the 50 ng ml–1 level) has been applied to the determination of low levels of lead in soft drinks, sediments and lichens. The results obtained show good agreement with certified or expected values.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.