Abstract

A method for the determination of arsenic (As) in seafood by inductively coupled plasma atomic emission spectrometry with continuous hydride generation is described. Several analytical parameters have been investigated and optimised. The analytical features of the method (recovery, precision, accuracy and limit of detection) were calculated. Practical detection limit of 3.6 μg/kg fresh weight for As has been reached. The precision of the method expressed as relative standard deviation (R.S.D.) was in the range of 2.7–3.7% and the recovery percentage ranged from 98.4 to 101.8%. The reliability of the developed method was checked by analysing several certified reference materials. A complete mineralization was obtained for arsenobetaine (AsB) containing reference material with a mixture of nitric and sulphuric acids followed by adding hydrogen peroxide in an open digestion system. This method can be applied to routine analysis without any risks of interferences.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.