Abstract
We present new HPLC methods for the quantitation in human plasma of two investigative metallotexaphyrin agents, motexafin gadolinium (Gd-Tex) and motexafin lutetium (Lu-Tex). Each assay uses: the other texaphyrin analogue as an internal standard; protein precipitation with acetonitrile:methanol (50:50, v/v); an ODS reversed-phase column; an isocratic mobile phase of 100 m M ammonium acetate, pH 4.3:acetonitrile:methanol (59:21:20, v/v/v); and absorbance detection at 470 nm. The Gd-Tex assay has a lower limit of quantitation (LLOQ) of 0.01 μ M and is linear between 0.01and 30 μ M. The Lu-Tex assay has an LLOQ of 0.1 μ M and is linear between 0.1 and 30 μ M. The assays are suited for in vivo preclinical studies and clinical trials because they require minimal amounts of plasma, are sensitive, and involve a 30-min run time. These assays are important tools for evaluating the potential of Gd-Tex and Lu-Tex as a radiation enhancer and photosensitizer, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.