Abstract

Abstract Strychnine was found measurable with high sensitivity by gas chromatography (GC) with surface ionization detection (SID) and using brucine as the internal standard. The calibration curve was linear in the range of 62.5 to 1000 pg, on column. The detection limit of strychnine was about 50 pg on column (2.5 ng/mL sample); the sensitivity was 70 times higher than that reported by high-performance liquid chromatography. A detailed procedure for extraction of strychnine from human whole blood and urine was established using Sep-Pak C18 cartridges; the recovery of strychnine and brucine, which had been spiked in both body fluids, was about 90%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.