Abstract

A sensitive and specific LC–MS–MS assay has been developed and validated for barnidipine (1-benzyl-3-pyrrolidinyl)methyl-2,6-dimethyl-4( m-nitrophenyl)-1,4-dihydropyridine-3,5-dicarboxylate). The assay involves a simple and rapid solid-phase extraction procedure. Sample analysis was on a Spherisorb S3ODS2 100 mm×2 mm I.D. column, with a Finnigan TSQ 7000 mass spectrometer, using an electrospray interface and selective reaction monitoring (SRM). The intra- and inter-day precision and accuracy, determined as the coefficient of variation and relative error, respectively, were 11.8% or less. The limit of quantitation was 0.03 ng/ml, and the calibration was linear between 0.03 and 3.0 ng/ml. The method has been used successfully for the measurement of over two thousand human plasma samples from pharmacokinetic clinical trials.