Abstract

Cortisol is one of the most important glucocorticoids involved in the regulation of human metabolism and physiological stress. Monitoring of levels of cortisol is of immense clinical benefit. In particular, salivary cortisol levels have been shown to correlate well with diurnal changes in cortisol levels in serum and have been used widely for monitoring of cortisol levels for diagnosis and prognosis purposes. We present a sensitive, fast, and robust quantitative liquid chromatography and tandem mass spectrometry (LC-MS/MS) assay for salivary cortisol in negative mode. This assay employs protein precipitation followed by reversed-phase liquid chromatographic separation, negative-mode electrospray ionization (ESI), and MS/MS detection. This assay has a total run time of 5.8minutes and a limit of quantification of 0.5ng/mL with a linear range up to 100ng/mL. No carryover was observed at 10μg/mL. This assay also incorporates the routine monitoring of prednisolone, a potential interferent to salivary cortisol.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.