Abstract

AbstractQuantitative µ‐XRF analysis based on standardless calibration is limited by the lack of information on the shape of the excitation spectrum resulting from using capillary lenses. The measurement of radiation scattered from a sample was used in combination with Monte Carlo simulation of radiation transport to estimate the energy spectral distribution of the excitation radiation. Further, a standardless calibration based on the fundamental parameter method implemented in the IAEA‐QXAS software package was carried out and verified for glass and fused ore certified reference materials. The accuracy and repeatability achieved are reported. Copyright © 2004 John Wiley & Sons, Ltd.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.