Abstract

Numerous studies have examined the histology of metal‐on‐metal hip tissues for evidence of a dose response to metal wear but have often reported inconclusive or contradictory findings. The aim of the present study was to address these discrepancies using multiple histological scoring methods to characterize the tissue features of one large group of revised metal‐on‐metal total hips. Periprosthetic tissues from 165 metal‐on‐metal hip revisions were examined for features of aseptic lymphocytic vasculitis associated lesions (ALVAL) as rated using two scoring systems as well as rankings for macrophage and lymphocyte numbers, intracellular wear debris and necrosis. Correlations between histological features and clinical variables including gender and time to revision and implant variables including articular surface wear volume or visual taper corrosion scores were examined. Both ALVAL scores reflected the macrophage dominated histology with average scores of 5.9/10 and 1.5/3. There was a statistically significant correlation between the original ALVAL score and wear rate per year (correlation coefficient = 0.17, p = .05) and a moderate correlation between the number of macrophages and wear particles and wear volume. There was no statistically significant correlation between wear and any other feature including lymphocytic inflammation or necrosis. Strong correlations between combined cup and ball wear volume and histological characteristics were not observed, although the number of macrophages was more closely correlated with wear than lymphocytes or necrosis.