Self-shielding in spatially distributed samples can be very significant when detecting photons below 100keV. A correction method has been developed for cylindrical samples, typically used in measurements of natural radioactivity with well-type detectors, as is the example of sediment dating. The method calculates the probability of photons to escape the sample, using Monte Carlo techniques with a program written in C. The effects of self-shielding on the angular distribution of photons have been indirectly analyzed as a function of sample geometry, given a fixed detector geometry. The results given by the program have also been used to provide a straightforward way of calculating the self-shielding factors, with uncertainties. Required inputs are the attenuation coefficient of the sample, its radius, and its height. A procedure has also been specified in order to make good use of all the information. To check on the method, measurements of reference material of known activity have been compared with calculated values, obtaining very satisfactory results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.