Abstract
The acid-treated multi-walled carbon nanotubes (MWNTs), which were modified on the surface of gold electrode, offers substantial improvements in voltammetric sensitivity and selectivity towards the determination of dopamine (DA). It can inhibit the voltammetric response of ascorbic acid (AA) while the redox reaction of dopamine is promoted. When a differential pulse voltammetric (DPV) technique was used, the peak separation between DAs and AAs was 244 mV. Based on this, a selective method could be constructed to detect DA in the presence of 1000 times higher concentration of AA. The effect of various experimental parameters on the voltammetric response of dopamine was investigated. Under the chosen conditions, the peak currents are correspondent linearly to the concentrations of DA in the range of 5×10 −7∼4×10 −4 mol L −1 with a limit of detection of 2×10 −7 mol L −1. The proposed method can be applied to detect DA in real samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.