Abstract

The most important steps in the validation of high-performance liquid chromatographic (HPLC) methods are discussed. The establishment of system suitability data and the assessment of peak purity are demonstrated on the example of bisquaternary amino steroids. For the recognition of incomplete resolution of adjacent peak pairs, the absorbance-ratio method in which the ratio of absorbances at two preselected wavelengths are plotted as a function of time in combination with the separation of sample components subjected to various chemical and physico-chemical treatments (stress conditions) is applied. The separation power and performance of the HPLC systems are characterized by the system resolution (SR) and system selectivity (SS). The special demands of stability-indicating methods are summarized.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.