Abstract

In the determination of nine β-agonists including salbutamol, terbutaline, cimaterol, fenoterol, clorprenaline, ractopamine, tulobuterol, clenbuterol and penbuterol in porcine, bovine, lamb and chicken muscle, liver and urine samples with LC-MS/MS, calibration curves prepared in solvent (SC) were compared with those prepared in each matrix (MC) for all analytes. Significant differences (P < 0.05) between SC and each MC for most analytes indicated the existence of matrix effects and the necessity of using MC for quantitation to compensate MEs. Then MC in each muscle, liver and urine sample was compared with the select potential representative matrix, followed by validating the recoveries of nine analytes calculated through the MC in the potential representative matrix and MC in their corresponding matrices, respectively. The results suggested that porcine muscle could be selected as a representative matrix to calibrate β-agonist residues in bovine, lamb and chicken muscle samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.