Abstract

In this paper, we present a fuzzy rule-based system for the automatic detection of seizures in the intracranial EEG (IEEG) recordings. A total of 302.7 hours of the IEEG with 78 seizures, recorded from 21 patients aged between 10 and 47 years were used for the evaluation of the system. After preprocessing, temporal, spectral, and complexity features were extracted from the segmented IEEGs. The results were thresholded using the statistics of a reference window and integrated spatio-temporally using a fuzzy rule-based decision making system. The system yielded a sensitivity of 98.7%, a false detection rate of 0.27/h, and an average detection latency of 11 s. The results from the automatic system correlate well with the visual analysis of the seizures by the expert. This system may serve as a good seizure detection tool for monitoring long-term IEEG with relatively high sensitivity and low false detection rate.