Abstract

Optical fibers have been proposed as dosimeters in both diagnostic and radiotherapy applications. A commercial germanium (Ge)-doped silica fiber with a 50 μm core diameter which showed good thermoluminescence (TL) properties was selected for this study. The radiation sources used were a high dose rate brachytherapy iridium-192, MV photon and MeV electron beams from a linear accelerator. The coating of the fiber was chemically removed and then annealed at 400 °C for 1 h prior to irradiation. After irradiation, the fiber was read on a Harshaw Model 3500 TLD reader. The optical fiber had one well-defined glow peak at 327 ± 2 °C at all the radiotherapy energies. The dose response was linear within the clinical relevant dose for all these energies. Reproducibility was mainly within 4–6% (one standard deviation) for high energy photons and electrons. The fiber was found to be energy independent within the MV photon energy range. At room temperature the fading up until 1 month was around 6% which was within the 6% uncertainty of the sensitivity calibration of the fiber. Re-using the fiber four times did not significantly alter the sensitivity factor. The optical fiber was found to be dose rate as well as angular independent. Central axis depth dose curves of both 10 MV photons and 12 MeV electrons using the fiber showed relatively good agreement to standard depth dose curves in water within 4%. The Ge-doped fiber is a promising TL dosimeter but improvements have to be made to reduce the reproducibility within 3% for high energy photons and electrons.