Abstract

The basics of second-order neutron optics for bent-crystal three-axis spectrometers are presented. Energy-transfer resolution for quasielastic scattering is examined in detail. Experimental data illustrating the effect of second-order aberrations are presented. The resolution of energy transfer at 5 meV neutron energy with bent monochromators made from commercial thin silicon wafers was improved to 8 µeV from a previous record of 16 µeV.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.