Abstract
Quasielastic scattering measurements with commercial silicon wafers as focusing monochromator and analyzer in a three-axis spectrometer showed energy-transfer resolutions in the 10–100 μeV range (fwhm). Resolutions were better and intensities higher than in a conventional arrangement with Soller collimators and pyrolytic graphite (PG) monochromator and analyzer. Resolution remained high when extended-plate samples were used in focusing orientation. At cold sources, this technique would give 10–20 μeV resolutions at neutron energies near the peak of the spectrum. Projections of resolution ellipsoids were determined by diffraction from powder samples. The orientations of the ellipsoids are controlled by horizontal curvatures and can be rotated 90° for Q-resolution. A monochromator unit with remote control of horizontal curvature and fixed vertical curvature set to spatial focusing at the sample position was also tested. The strong vertical focusing gave a significant gain in integrated intensities, but worsened the resolution because of second-order aberrations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.