Abstract

A highly selective and sensitive second-derivative spectrophotometric determination of iron is based on the extraction of the ferroin—perchlorate ion-association complex into mesityl oxide. A linear calibration graph is obtained for iron in the range 0.5–50 μg in 100 ml of original aqueous phase, with a detection limit of 2 ng/ml. The method is precise and reliable and has been applied to the determination of iron in high-purity rare-earth oxides.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.