Abstract

Sixteen embalmed hemipelves were used to determine the optimal acetabular screw placement to provide maximal screw pull-out strength in unicortical and bicortical screw fixation. The anterior column, superior ilium, posterior column, ischium, and pubis regions of the pelvis were tested using 6.5-mm titanium alloy screws and a hydraulic servo-controlled 1321 Instron testing machine. Force vs displacement data were acquired. Bicortical fixation was stronger than unicortical fixation in the four zones compared. This difference was significant in the superior ilium, posterior column, and ischium. The anterior column could not accept unicortical screws due to inadequate bone depth, which ranged between only 6 mm and 10 mm. Bicortical fixation was significantly greater in the superior ilium, posterior column, and ischium than in the anterior column or pubis. Unicortical fixation was greatest in the superior ilium. This information may aid decisions concerning the positioning of screws to augment acetabular component fixation.